Market Forge Sterilizer - Timer
Details
How to replace the timer:
1
2. Remove top cover, . Disconnect supply power,
3. Undo the timer knob on the front panel with Allen key.
4. Remove 3 small screws from the front of the timer behind the knob,
5. Remove the timer, replace the wires in same order on the new
timer and follow the other steps in reverse order.
6. There is a diagram, located on the bottom of left side, there is a small window, which is the wire connection box. On the inside of this small window cover, there is a wire diagram.
7. Click here for Timer Calibration Instructions
1
2. Remove top cover, . Disconnect supply power,
3. Undo the timer knob on the front panel with Allen key.
4. Remove 3 small screws from the front of the timer behind the knob,
5. Remove the timer, replace the wires in same order on the new
timer and follow the other steps in reverse order.
6. There is a diagram, located on the bottom of left side, there is a small window, which is the wire connection box. On the inside of this small window cover, there is a wire diagram.
7. Click here for Timer Calibration Instructions